A calibration report, a catalog accuracy statement and a project acceptance decision answer different questions. Buyers need to know what was measured, the uncertainty associated with the result and the rule used to decide whether the instrument is acceptable for its intended role. Treating these documents as interchangeable creates avoidable ambiguity at handover.
This guide describes procurement review rather than a calibration method or a conformity certificate. It does not set a universal tolerance, uncertainty ratio or guard band. The project owner must agree the acceptance rule and any required evidence with the calibration provider before testing.

Application Requirements for System Integrators
An engineering contractor may need evidence for a new transmitter before commissioning. An OEM may instead need a service comparison to investigate a disputed reading. These activities have different scope: an informal field comparison does not automatically become a traceable calibration or a formal acceptance decision.
Define whether the acceptance concerns the probe alone, the transmitter output or the complete receiving chain. A test of the local display cannot establish the behavior of an untested analog input or gateway conversion. The purchase specification should identify the boundary and the required reporting format.
Nexisense Product Reference
Nexisense publishes the HG803 as a temperature and humidity measurement product. The linked product reference identifies the catalog configuration and the table separates its published information from project checks. Obtain the order-specific drawing and specification before procurement. Related product photos are references, not proof that different probes or housings are interchangeable.
Product specification (English): HG803
Product Parameters and Procurement Checks
| Parameter | Published Reference | Engineering Check |
|---|---|---|
| Model | HG803 | Match report identity to the installed assembly. |
| Supply | DC 10-30V | Define whether the powered transmitter is in scope. |
| Outputs | 4-20mA / 0-5V / 0-10V / RS485 | State which output and receiving path are tested. |
| Protocol | Modbus-RTU | Confirm the tested digital representation where relevant. |
| Humidity range | 0-100% RH | Check that the test points cover intended use. |
Selection Guide
Request the instrument identity, tested configuration, test points, conditions, reference information and reported uncertainty. Where a correction is applied, clarify how it is recorded and whether the acceptance refers to results before or after adjustment. Do not equate a stated display resolution with measurement uncertainty.
Agree how results near an acceptance boundary will be handled. A decision rule may account for uncertainty, but its choice is a project agreement rather than a property of the sensor model. Record who is authorized to accept an indeterminate or nonconforming result.

Integration, Installation and Commissioning
Check that the report identifier matches the installed device and probe combination. Review the tested range against the intended operating points. A report covering a narrow stable condition does not establish performance over untested temperatures, humidity conditions or contamination exposure.
For a field comparison, stabilize the reference and device in representative conditions and record their positions. An observed difference includes environmental gradients and both instruments; it cannot automatically be assigned to the transmitter. Escalate a disputed result rather than applying an undocumented offset.
Preserve the acceptance rule, report revision and configuration used for the decision. If a probe or signal path changes, review whether the earlier evidence still covers the installed chain. Keep calibration status distinct from alarm settings and from the process owner acceptance of product quality.
Illustrative Engineering Application
Illustrative engineering example: an OEM receives a humidity transmitter and requests calibration evidence for its intended operating region. The quality team agrees the report scope and decision rule before release. A result near the boundary is referred for review rather than being accepted solely from a catalog accuracy figure. No actual calibration result or conformity outcome is asserted.

Why Consider Nexisense
Nexisense provides published product references that can support a structured technical discussion. Submit the operating conditions, interface needs and acceptance scope for configuration review. Base purchasing approval on confirmed documentation and the required project evidence, not on photographs or an assumed capability shared by every variant.
Frequently Asked Questions
Q1. Is catalog accuracy the same as calibration uncertainty?
No. They describe different aspects of measurement information. Review the reported calibration result and uncertainty in the context of the project requirement rather than substituting one term for the other.
Q2. Does a handheld comparison constitute calibration?
Not automatically. Its status depends on the method, references, conditions and reporting scope. A useful service comparison should not be presented as formal traceability evidence without the required basis.
Q3. What is a decision rule?
It is the agreed basis for interpreting results against an acceptance requirement, including how uncertainty is handled where applicable. The project should specify it before the decision is made.
Q4. Can display resolution establish uncertainty?
No. Display increments do not describe all measurement influences. Request the appropriate uncertainty information from the calibration provider and review the scope of the reported result.
Q5. Should pre-adjustment results be retained?
Where required by the agreed scope, they help explain the instrument condition before intervention. Clarify reporting expectations before adjustment rather than discovering missing evidence during acceptance.
Q6. Does a probe report cover the entire PLC chain?
Not necessarily. Define the tested boundary explicitly. Signal conversion, receiving input and gateway processing require appropriate verification if the acceptance concerns the complete installed chain.
Q7. Can an offset settle a disputed comparison?
An offset alone does not explain the cause of disagreement. Review conditions, references and the approved method before changing an instrument that is used for control or acceptance.
Q8. Who accepts a borderline result?
The project quality plan should identify the responsible authority and agreed rule. This article does not assign an automatic pass, fail or universal guard band.
Summary
Make humidity acceptance evidence explicit: tested identity, scope, conditions, reported uncertainty and an agreed decision rule. Keep field comparisons, calibration reports and process acceptance distinct. Preserve the decision record through subsequent probe, output and configuration changes.
Further Reading: Related product specification (English); Temperature and humidity product range (English)